Analysis of Diagnostic Reference Level (DRL) and Effective Dose Values based on Age and Gender in Head CT Scan Examinations

Authors

DOI:

https://doi.org/10.24843/BF.2027.v28.i01.p10

Keywords:

Effective dose, Diagnostic Reference Level (DRL), CT scan, CTDIvol, DLP

Abstract

Computed Tomography (CT) examinations involve relatively high radiation doses, making radiation dose optimization essential. This study aimed to evaluate local Diagnostic Reference Level (DRL) values and Effective Dose in non-contrast head CT examinations based on patient age and gender. A total of 270 patient datasets were retrospectively analyzed using a 256-slice Philips iCT scanner operated at 120 kV and 400 mAs. Local DRL values were determined using the third quartile (Q3) of CTDIvol and DLP distributions. The results showed that DLP values were slightly higher in the ≥15 years group (1234.40 mGy·cm) than in the 5–14 years group (1212.68 mGy·cm). Male patients also demonstrated higher DLP values (1243.30 mGy·cm) compared to female patients (1223.10 mGy·cm), while CTDIvol remained constant at 53.34 mGy across all groups. The mean Effective Dose for adult patients was 2.15 ± 0.18 mSv in males and 2.12 ± 0.17 mSv in females. Statistical analysis showed no significant difference in DLP values based on age (p = 0.09), but a significant difference based on gender (p = 0.015). No significant difference was found in Effective Dose between adult male and female patients (p = 0.081). All dose parameters were below the national DRL limits established by BAPETEN, indicating that the non-contrast head CT protocol at Hospital X complies with radiation protection optimization principles and national safety standards.

Downloads

Published

2026-11-20

Most read articles by the same author(s)